Atlas FDA

3M Attest Rapid Readout 1292 Biological Indicators (Steam) Pack 70-2006-2052-7, Lot 2006-21, in boxes of 50 per box

FDA device recall Z-0720-04 · posted 2004-07-20 · Terminated

Recall details

Recalling firm
3M Company / Medical Division
Reason for recall
The 3M Attest 1292 Rapid Readout Biological Indicators for steam sterilization may contain ampoules which may result in a false negative readout at 3 hours.
Action taken
A Product Recall Notice was sent on February 27, 2004 to consignees instructing them to discontinue use of the recalled product and return products to 3M.
Root cause
Other
Status
Terminated
Product code
FRC
Quantity affected
90,000 ampoules
Distribution
Nationwide throughout the United States and worldwide.
Date initiated
2004-02-27
Date posted
2004-07-20
Date terminated
2006-12-28
Location
St Paul, MN