3M Attest Auto-reader. Model numbers 390, 390G,and 490. These devices incubate the biological indicators (BIs) to determ
FDA device recall Z-2002-2014 · posted 2014-07-07 · Terminated
Recall details
- Recalling firm
- 3M Company - Health Care Business
- Reason for recall
- Labeling on units shipped prior to May 16th, 2014 did not contain the statement "This product contains dry natural rubber" as required by the United States Food and Drug Administration (FDA) for medical devices. The non-slip pads at the bottom of the unit contain dry natural rubber. One incident of an allergic reaction in a sensitized individual has been reported.
- Action taken
- Consignees were sent on 6/9/2014 a 3M "Urgent Medical Devices Correction" letter dated May 21, 2014. The letter described the problem and the product involved in the recall. The letter described what action is 3M taking and what action was required of the consignees. Consignees were sent an adhesive label and a response card. They requested consignees to adhere the label to a portion of the plastic covering each unit in a location where it is visible to the end-user. After that, they are requested to fill out and return the response card.
- Root cause
- Labeling Change Control
- Status
- Terminated
- Product code
- FRC
- Quantity affected
- 5343 (2621 USA, 2722 OUS)
- Distribution
- Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the countries of AUSTRALIA, CANADA, CHILE , COLOMBIA, COSTA RICA, SOUTH AFRICA, BRAZIL, GERMANY, ECUADOR, DUBAI, JAPAN, HONG KO
- Date initiated
- 2014-06-09
- Date posted
- 2014-07-07
- Date terminated
- 2015-06-03
- Location
- Saint Paul, MN
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| 3M ATTEST(TM) AUTO-READER (K123546) | 3M Company | 2013-03-18 |
| 3M ATTEST 1491 SUPER RAPID READOUT BIOLOGICAL INDICATOR; 3M ATTEST 490 AUTO-READER (K103277) | 3M Company-3M Health Care | 2011-04-19 |