Atlas FDA

3M Attest 1264P Biological Indicators, in boxes of 25 per box

FDA device recall Z-0237-04 · posted 2003-12-18 · Terminated

Recall details

Recalling firm
3m Health Care
Reason for recall
The 3M Attest 1264/1264P Biological Indicators of ETO sterilization contain a microbiological contaminant which can affect the performance of the positive control and the indicators to some limited extent. The color on a positive control may revert to negative after 24 hours of incubation.
Action taken
The firm conducted a field correction which began on 10/24/03. They provided revised labeling for the product. Letters, dated October 24, 2003, were sent to consignees. The letters had a description of the problem, instructions for doing the relabeling in the field, and recommendations for users of the product.
Root cause
Other
Status
Terminated
Product code
FRC
Quantity affected
1844 boxes of 25 biological indicators per box
Distribution
The product was distributed nationwide in the United States and worldwide.
Date initiated
2003-10-24
Date posted
2003-12-18
Date terminated
2005-02-09
Location
Saint Paul, MN