300 microliter CO-RE tips for use with Microlab STAR series instruments, Part Number 235902, 5,760 tips per case, manufa

FDA device recall Z-1203-2009 · posted 2009-04-17 · Terminated

Recall details

Recalling firm
Hamilton Co
Reason for recall
Some tips may be occluded and are unable to aspirate or dispense liquid.
Action taken
An "Urgent: Product Recall Letter" dated December 1, 2008 was issued to consignees via certified mail and e-mail. The letter described the problem and potential effects of use. Consignees were instructed to complete the enclosed "Hamilton Company Product Recall Questionnaire" and return it via e-mail or fax (775-856-7259) to the Hamilton Company. After receipt of the completed questionnaire, the firm's Customer Service Department will contact consignees to provide Returned Material Authorization information for returning affected product. Direct questions to the Hamilton Company at 775-858-3000, ext. 236.
Root cause
Process control
Status
Terminated
Product code
JTC
Quantity affected
28 cases
Distribution
Nationwide Distribution -- MA, CA, MN, CT, and NC.
Date initiated
2008-12-01
Date posted
2009-04-17
Date terminated
2009-07-21
Location
Reno, NV