300 Microliter CO-RE Tips for use with Microlab STAR, manufactured by Hamilton Company, Reno, NV. Product is a disposabl

FDA device recall Z-1999-2009 · posted 2009-09-09 · Terminated

Recall details

Recalling firm
Hamilton Co
Reason for recall
Product Defect-- some tips from the mold cavity 28 are too short and do not always aspirate or dispense liquid correctly.
Action taken
The firm initiated the recall on April 10, 2009, and issued notification to consignees via certified mail and e-mail. Customers were instructed to return the affected product to the Hamilton Company. A fax-back form was supplied with each notification. Direct questions about the recall to the Hamilton Company by calling 1-775-858-3000, extension 236.
Root cause
Process control
Status
Terminated
Product code
JTC
Quantity affected
20 cases of Part 235950 and 27 cases of Part 235902.
Distribution
Worldwide Distribution -- US (states of DE, NJ, PA, MA and CA), New Zealand and Japan.
Date initiated
2009-04-09
Date posted
2009-09-09
Date terminated
2010-01-04
Location
Reno, NV