3.2mm Drill Bit. This is a component within the PediLoc Locking Plate System utilized to pre-drill a hole in bone prior

FDA device recall Z-1941-2013 · posted 2013-08-09 · Terminated

Recall details

Recalling firm
OrthoPediatrics Corp
Reason for recall
OrthoPediatrics has initiated a voluntary recall of Part # 01-1050-002, 2.5mm Drill Bit, Lot #91300335. Only drill bits that are in their original packaging are being recalled. The description on the package of this drill bit has a wrong size in the description (3.2mm). The recall affects only packaged devices still in original packaging due to misbranding of the product. The drill bit itself has
Action taken
On 7/24/13, OrthoPediatrics notified their consignees of this URGENT MEDICAL DEVICE RECALL via phone call. A follow-up email message was then sent with the recall notification, and then a hard copy of the notice was sent via certified letter. The notification addresses the affected product, recall issue, reply form instructions, and directs consignees to return affected product to OrthoPediatrics via a Fed-Ex return prepaid pouch. The firm directs consignees to forward the notification and reply form if the product was further distributed.
Root cause
Process design
Status
Terminated
Product code
HWC
Quantity affected
68
Distribution
Nationwide distribution, including the states of MO, MI, WA, FL, NY, TN, CO, AZ, VA, TX, CA, MN, LA, NC, and UT.
Date initiated
2013-07-24
Date posted
2013-08-09
Date terminated
2013-08-26
Location
Warsaw, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ORTHOPEDIATRICS PEDILOC LOCKING PLATE SYSTEM (K083286)OrthoPediatrics Corp.2009-01-30