2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-9

FDA device recall Z-2215-2024 · posted 2024-06-27 · Open, Classified

Recall details

Recalling firm
Thoratec Switzerland GMBH
Reason for recall
Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system.
Action taken
On 5/22/24, correction notice distribution started. Customers were asked to do the following: If the console shuts down without alarms or alerts present, follow Instruction for Use (IFU) for Console and Motor malfunction as described in the Section Emergency / Troubleshooting , and switch to a backup console, which must be prepared as required backup components in the immediate vicinity. As stated in the IFU these backup systems should have a battery charge sufficient for at least one hour of operation, be connected to wall power, and be powered off. Note: When the backup console is put into use, the previously used console can be restarted and set up as a backup console. Per the IFU and current practice, a full console backup system is always required to be available in the immediate vicinity of each patient whenever the system is used. Firm is in the process of implementing an additional in-process test by Q3 of 2024, which will ensure that all systems which will be distributed are confirmed to withstand a surge up to 2.0 kV. Distribute this notice to those who need to be aware within your institution and to any organization where potentially affected devices may have been transferred in the event that devices were transferred elsewhere. Complete and return the acknowledgement form via email to CentriMagNotices@abbott.com Should you have any questions about this communication, please contact firm Customer Service at 1 (800) 456-1477
Root cause
Process control
Status
Open, Classified
Product code
DSQ
Quantity affected
3,425
Distribution
Worldwide - US Nationwide distribution included in the states of CA, PA, FL, IL, NY, AR, GA, WI, AZ, TN, TX, NC, MT, VA, MA, NJ, NE, WV, ME, OK, CO, OH, KC, WA, IN, CT, MI, OR, ID, MO, SC, KY, LA, NM,
Date initiated
2024-05-22
Date posted
2024-06-27
Location
Zurich, Switzerland