2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600
FDA device recall Z-1962-2025 · posted 2025-06-11 · Open, Classified
Recall details
- Recalling firm
- Baxter Healthcare Corporation
- Reason for recall
- The replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to excessive heat and user burns.
- Action taken
- Baxter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 05/09/2025 via USPS First Class mail. The notice explained the issue, hazard involved, and requested to cease all use of the device and return it to Baxter. Please contact Baxter Technical Support to arrange for return and replacement of your affected battery or Welch Allyn PocketScope handle at 800-535-6663 between the hours of 8:00 am and 8:00 pm Eastern Time, Monday through Friday. Press option 2, then select option 1 or 2 for language preference, then press option 1 for physical exam, and then press 3 for physical exam products. Forward a copy of the notice to any other departments with the institution that may use the affected product. Distributors were directed to notify their consignees.
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- ERA
- Quantity affected
- 277 units
- Distribution
- US Nationwide distribution.
- Date initiated
- 2025-05-09
- Date posted
- 2025-06-11
- Location
- Deerfield, IL
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.