2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with

FDA device recall Z-2738-2015 · posted 2015-09-04 · Terminated

Recall details

Recalling firm
Trilliant Surgical Ltd.
Reason for recall
Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.
Action taken
The firm sent out a written notification of the recall on 7/31/2015.
Root cause
Component design/selection
Status
Terminated
Product code
HRS
Quantity affected
108 units
Distribution
NY, OH, CA, SC, NJ, AZ, VA, IN, FL, MN, OK, AR, MO, WI, GA, SD, IA, IL, TX
Date initiated
2015-07-31
Date posted
2015-09-04
Date terminated
2017-01-26
Location
Houston, TX