18 Fr 6mm Sft Flow Angled Wire The Sarns Aortic Cannula is indicated for use in perfusion of the ascending aorta during

FDA device recall Z-0459-2013 · posted 2012-12-03 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic Cannula. The plastic flash has the potential to detach and contribute to an adverse patient effect. While this potential failure has not been reported from the field as an observed defect, it was determined that the product does not meet specifications and this condition could result in an unacceptabl
Action taken
TERUMO sent an URGENT MEDICAL DEVICE RECALL letter dated September 20, 2012, to all affected customers. The letter identified the product, the problem, and the aciton to be taken by the customer. Customers were instructed to remove this cannulae from circulation and return them for credit. Customers with questions should call 1-800-521-2818. For questions regarding this recall call 734-741-6173.
Root cause
Process design
Status
Terminated
Product code
DWF
Quantity affected
5760 total
Distribution
Worldwide Distribution - USA including AL , CA , CO , DC , DE , FL , GA , IL , IN , KS , KY , LA , MA , MI , MO , MS , MT , NC , ND , NE , NH , NY , OH , PA , TN , TX , VA , and WI. Internationally to
Date initiated
2012-11-21
Date posted
2012-12-03
Date terminated
2013-04-23
Location
Ann Arbor, MI