15 L Cycler Drainage Bag Product Usage: For use with Baxter Cycler Tubing sets as a peritoneal dialysis accessory.
FDA device recall Z-0080-2015 · posted 2014-10-17 · Terminated
Recall details
- Recalling firm
- Baxter Healthcare Corp
- Reason for recall
- The large tube clamp used to close the drain tube on the 15 L Cycler Drainage Bag may not fully close the drain tube and may result in leakage.
- Action taken
- Baxter sent an Urgent Product Recall letter dated May 30, 2014 to all affected customers. The letter identified the affected product, problem and actions to be taken. The recalling firm is instructing their customers to remove all affected products from their facility and quarantine the product subject to the recall. Customers are instructed to return the affected product to Baxter and receive a credit or replacement. For questions contact the Center for One Baxter at 1-800-422-9837.
- Root cause
- Component change control
- Status
- Terminated
- Product code
- KDJ
- Quantity affected
- 3,085,080
- Distribution
- Worldwide Distribution - US Nationwide (including Puerto Rico) and countries of, Canada, Australia, New Zealand, Thailand, Malaysia, Singapore, Korea, Japan, Philippines, Columbia, Ecuador, Austria, B
- Date initiated
- 2013-10-17
- Date posted
- 2014-10-17
- Date terminated
- 2016-01-04
- Location
- Deerfield, IL