100/120V AC, Advanced Perfusion System Platform (APS) The Terumo Advanced Perfusion System 1 is indicated for use for up

FDA device recall Z-1687-2015 · posted 2015-05-28 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
Terumo CVS is implementing field correction activities to address the identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late 2015: - Implementing design changes and upgrades to improve the reliability of power switching and battery backup. - Adding a switch protector to the On/Off power switch to preven
Action taken
On 5/1/2015 Terumo issued an URGENT MEDICAL DEVICE CORRECTION letter to users explaining that Terumo has identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late 2015: - Implementing design changes and upgrades to improve the reliability of power switching and battery backup. - Adding a switch protector to the On/Off power switch to prevent inadvertent shut off. - Replacing the On/Off power switch with a new design to prevent switch failure. A Terumo Field Service Representative will contact users to schedule the field correction activities once the activities are ready to be implemented. CUSTOMER INSTRUCTIONS: Review this Medical Device Correction and assure that all users have received notice of this issue. - Complete and return the enclosed Customer Response Form. - Terumo CVS will contact users to schedule the field correction activities. We encourage you to contact us with any questions or concerns: Terumo CVS Customer Service 1-800-521-2818 Fax 1-734-741-6149 Customer Service Hours: Monday Friday, 8 a.m. 6 p.m. ET
Root cause
Device Design
Status
Terminated
Product code
DTQ
Quantity affected
1713 units total
Distribution
Worldwide Distribution-US (nationwide) including DC and the states of AL , AR , AZ , CA , CO , CT , DE , FL , GA , HI , IA , ID , IL , IN , KY , LA , MA , MD , ME , MI , MN , MO , MS , MT , NC , ND ,
Date initiated
2015-05-01
Date posted
2015-05-28
Date terminated
2018-09-28