.045 X 6 ST Guide Wire, Part number 45-0029-S, Sterile 10 pack of WS-1106ST and has a 6 inch length and 0.045 inch diame
FDA device recall Z-1483-2017 · posted 2017-03-16 · Terminated
Recall details
- Recalling firm
- Acumed LLC
- Reason for recall
- Guide wires sized as .045 X 6 were mis-packaged as .054 X 7 guide wires and those sized as .054 X 7 were mis- packaged as .045 X 6 guide wires. There is a possibility that using the incorrect guide wire size can cause health hazards, such as mal or nonunion, need for revision surgery, and histological response (if wire breaks).
- Action taken
- Acumed sent an Urgent Medical Device Recall letter dated 02/16/2017, to all affected customers on 02/17/2017. Consignees were instructed to follow special instructions and send survey to RecallNotification@acumed.net. Domestic customers with questions or need to return products to Acumed, please contact Acumed Agent Inventory via email AgentInventory@acumed.net to obtain a Return Material Authorization (RMA) number or contact your agent directly. International customers with questions or need to return products to Acumed, please contact Acumed Business Service via email at BusinessServices@acumed.net or call to obtain a Return Material Authorization (RMA) number. Customers with questions about recall can call 877-627-9957, choose Option 4 (US and Canada), or call 001-503-627-9957, choose option 4 (outside of US and Canada).
- Root cause
- Process control
- Status
- Terminated
- Product code
- LXH
- Quantity affected
- 24 units of .045 X 6 ST Guide Wire and 25 units of .054 X 7 ST Guide Wire
- Distribution
- Worldwide Distribution in Great Britain, Ireland, and Japan only.
- Date initiated
- 2017-02-17
- Date posted
- 2017-03-16
- Date terminated
- 2017-05-19
- Location
- Hillsboro, OR