00597206126 Patella ALL POLY PAT COMP MICRO 26DIA, Rx, Sterile; 00597206129 Patella ALL POLY PAT COMP MICRO 29DIA, Rx, S
FDA device recall Z-2015-2013 · posted 2013-08-22 · Terminated
Recall details
- Recalling firm
- Zimmer, Inc.
- Reason for recall
- Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
- Action taken
- Beginning June 7, 2013 all distributors will be sent Urgent Device Recall notices via electronic mail, with distributors, hospital risk managers with affected inventory also receiving the notice via courier. The Notice outlined the products recalled; the reason for the recall, possible adverse events related to the recall and directed consignees to locate, quarantine and count all affected product recording the information on the Inventory Return Form acknowledging that recall directions were followed. Further questions should be directed to 1-877-946-2761. Affected product is to be returned to Zimmer Product Service Department , 1777 West Center St. Warsaw, IN 46580 or internation return at Zimmer International Logistics GmbH, Max - Immelmann Allee 12 79427 Eschbach Germany. " Communications outside of the United States will occur approximately two weeks after the United States communications.
- Root cause
- Labeling design
- Status
- Terminated
- Product code
- JWH
- Quantity affected
- 192,355 all devices
- Distribution
- Worldwide Distribution-USA (nationwide) and the country of Japan.
- Date initiated
- 2013-06-07
- Date posted
- 2013-08-22
- Date terminated
- 2014-12-15
- Location
- Warsaw, IN