00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter 130 mm Stem L
FDA device recall Z-1092-2014 · posted 2014-02-26 · Terminated
Recall details
- Recalling firm
- Zimmer, Inc.
- Reason for recall
- This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of the minimum required by specification in one of the three zones of the Trabecular Metal. The other two zones exceeded the minimum specification.
- Action taken
- Zimmer sent an Urgent Medical Device Recall dated January 10, 2014 to distributors via electronic mail. The hospital/surgeon that has implanted this manufacturing lot will be identified and notified by the distributor. The letter identified the affected product, problem and actions to be taken. Distributors were instructed to locate and remove the affected product in their territory, as well as identifying and notifying the hospital/surgeon that has implanted this manufacturing lot as applicable. The distributor will return on-hand affected products to Zimmer and ensure all of their products are accounted for using the return form provided in the letter. Distributors and hospital consignees are to complete and return the inventory return form and acknowledgement form indicating which hospital/surgeon has implanted the product and to which they have provided notice to. Email a copy of the completed return form to CorporateQuality.PostMarket@Zimmer.comCommunications outside of the United States will occur approximately two weeks after the United States communications. For questions call 1-877-946-2761
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- KWT
- Quantity affected
- 10 units
- Distribution
- Distribution USA in the state of Illinois.
- Date initiated
- 2014-01-10
- Date posted
- 2014-02-26
- Date terminated
- 2014-09-11
- Location
- Warsaw, IN
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| ZIMMER TRABECULAR METAL HUMERAL STEM (K041549) | Zimmer, Inc. | 2004-09-01 |