Atlas FDA

0.9% Sodium Chloride: Brand Name: Product Name/Product Code: MAC MEDICAL: 10ML FILL SYRINGE/9210; NURSE ASSIST: 3ML IV F

FDA device recall Z-0511-2024 · posted 2023-12-20 · Open, Classified

Recall details

Recalling firm
Nurse Assist, LLC
Reason for recall
Sodium chloride irrigation USP, and sterile water for irrigation USP, that may be packaged inside kits, have the potential for a lack of sterility assurance, which could result in a nonsterile product, use of which could cause infection.
Action taken
On 11/6/23, the firm issued a press release. On 11/09/23, "URGENT: MEDICAL DEVICE RECALL" notices and response forms were emailed and mailed to direct customers and distributors, who were provided with a separate customer recall notice they could forward to their customers. Customers were asked to do the following: 1) Quarantine and then return affected devices. 2) If affected products were further distributed, send these customers the attached recall notice that asks customers to return affected devices to their distributors. 3) Distributors are asked to follow up with customers to ensure they received the recall notification and are taking appropriate action. 4) Distributors are asked to provide their Customer Shipment Report that lists all devices shipped to customers. The report should include the Shipment Date, Customer Name and Address, Part Number, Lot Number, and Quantity. This information is important to determine removal effectiveness. 5) Complete and return the response form via email to nurseassist@rqa-inc.com 6) If your company repacked these medical devices into convenience kits, please contact your local FDA recall coordinator to evaluate the need for a new recall. The firm is arranging for removal and replacement of the affected product. The recall notice sent by distributors to their customers also asks that the notice be forwarded to the applicable personnel in their organization or to any organization where the Affected Devices were transferred. Customers with questions can call 800-649-6800, between the hours of 8am and 4:30pm (CST), or email productremovalinfo@nurseassist.com
Root cause
Process control
Status
Open, Classified
Product code
NGT
Quantity affected
61,740 Cases
Distribution
US Nationwide Distribution: SC AZ, CA, FL, GA, HI, IA, IN, KS, KY, MA, ME, MN, MS, NC, ND, NE, NJ, NV, NY, OH, PA, RI, TX, WA, WI, AK, AL, AR, CO, CT, DC, ID, IL, LA, MD, MI, MO, MT, NH, NM, OK, OR, P
Date initiated
2023-11-06
Date posted
2023-12-20
Location
Haltom City, TX

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
0.9% Sodium Chloride I.V. Flush Syringe for Sterile Delivery, 1 syringe/pouch, 0.9% Sodium Chloride I.V. Flush Syringe for Sterile Delivery, 2 syringes/pouch (K201286)Nurse Assist, LLC2020-08-11
WOUND FLUSH, STERILE WATER & NORMAL SALINE (K083042)Nurse Assist, Inc.2008-12-18
PRIMARY CARE SOLUTIONS PREFILLED 10CC AND PREFILLED 30CC INFLATION SYRINGES WITH STERILE WATER CATALOG NUMBERS 1010 AND (K030813)Primary Care Solutions, Inc.2003-05-30