Atlas FDA

Vascade Vascular Closure System (VCS)

FDA premarket approval P120016/S029 · decided 2023-03-01

Approval details

PMA number
P120016
Supplement
S029
Supplement type
Normal 180 Day Track
Supplement reason
Change Design/Components/Specifications/Material
Trade name
Vascade Vascular Closure System (VCS)
Generic name
Device, hemostasis, vascular
Applicant
Cardiva Medical, Inc.
Date received
2022-09-02
Decision date
2023-03-01
Days to decision
180 days
Decision code
APPR
Product code
MGB
Advisory committee
Cardiovascular
Expedited review
No
Location
Santa Clara, CA
Statement
Approval for the following changes:1) Increase in disc deployment diameter from 23F to 25F;2) Increase in collagen mass from 12 mg to 19 mg;3) Modifications to certain device components to accommodate the new disc deployment diameter and collagen mass; and4) Modifications to the Instructions for Use to reflect the changes in device design.

Class III intelligence, same tools

PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.

510(k) clearances by this applicant

RecordFirmDate
CATALYST II AND III (K130124)Cardiva Medical, Inc.2013-03-14
CARDIVA CATALYST III SYSTEM WITH PROTAMINE SULFATE (K082930)Cardiva Medical, Inc.2009-02-13
BOOMERANG CATALYST/BOOMERANG CATALYST II SYSTEM (K072297)Cardiva Medical, Inc.2007-09-12
BOOMERANG WIRE /BOOMERANG PLUS WIRE SYSTEM, MODEL# B PLUS WIRE 58L (K070485)Cardiva Medical, Inc.2007-03-22
BOOMERANG CLOSUREWIRE VASCULAR CLOSURE SYSTEM (K051817)Cardiva Medical, Inc.2006-12-06
BOOMERANG WIRE (K061075)Cardiva Medical, Inc.2006-09-01
VASOSTASIS VASCULAR CLOSURE SYSTEM (K041486)Cardiva Medical, Inc.2004-10-22