THERAPY ABLATION CATHETER
FDA premarket approval P040014/S015 · decided 2011-01-11
Approval details
- PMA number
- P040014
- Supplement
- S015
- Supplement type
- Real-Time Process
- Supplement reason
- Change Design/Components/Specifications/Material
- Trade name
- THERAPY ABLATION CATHETER
- Generic name
- Cardiac ablation percutaneous catheter
- Applicant
- Irvine Biomedical, Inc.
- Date received
- 2010-11-16
- Decision date
- 2011-01-11
- Days to decision
- 56 days
- Decision code
- APPR
- Product code
- LPB
- Advisory committee
- Cardiovascular
- Expedited review
- No
- Location
- Irvine, CA
- Statement
- APPROVAL FOR A DESIGN MODIFICATION TO THE TIP ELECTRODE STEM OUTER DIAMETER (OD) SO THAT THE GAP BETWEEN THE TIP ELECTRODE OD AND TUBE INSIDE DIAMETER (ID) IS REDUCED TO -0.0015¿ TO +0.0015¿ FROM -0.003¿ TO +0.003¿ FOR CONSISTENT GAP INTERFERENCE.
Class III intelligence, same tools
PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.
510(k) clearances by this applicant
| Record | Firm | Date |
|---|---|---|
| INQUIRY H-CURVE TV STEERABLE DIAGNOSTIC CATHETER (K082061) | Irvine Biomedical, Inc. | 2008-08-19 |
| INQUIRY OPTIMA PLUS STEERABLE ELECTROPHYSIOLOGY CATHETER (K060757) | Irvine Biomedical, Inc. | 2006-04-19 |
| INQUIRY AFOCUS II DIAGNOSTIC CATHETER (K053582) | Irvine Biomedical, Inc. | 2006-03-21 |
| INQUIRY AFOCUS, INQUIRY AFOCUS II AND INQUIRY OPTIMA STEERABLE ELECTROPHYSIOLOGY CATHETER (K042775) | Irvine Biomedical, Inc. | 2004-11-04 |
| AFOCUS ELECTROPYSIOLOGY CATHETER SYSTEM, MODELS 81550, 81567, 81568, 81569 (K010471) | Irvine Biomedical, Inc. | 2001-06-06 |
| LUMA-CATH STEERABLE EP CATHETERS (K991878) | Irvine Biomedical, Inc. | 1999-12-10 |
| LUMA-CATH FIXED CURVE EP CATHETERS (K991025) | Irvine Biomedical, Inc. | 1999-10-15 |
| ORTHOGONAL EP CATHETERS (K990958) | Irvine Biomedical, Inc. | 1999-06-03 |