TECNIS Odyssey IOL with TECNIS SIMPLICITY Delivery System (DRN00V); TECNIS Odyssey Toric II IOLs with TECNIS SIMPLICITY
FDA premarket approval P980040/S174 · decided 2025-09-11
Approval details
- PMA number
- P980040
- Supplement
- S174
- Supplement type
- Real-Time Process
- Supplement reason
- Change Design/Components/Specifications/Material
- Trade name
- TECNIS Odyssey IOL with TECNIS SIMPLICITY Delivery System (DRN00V); TECNIS Odyssey Toric II IOLs with TECNIS SIMPLICITY
- Applicant
- Johnson & Johnson Surgical Vision, Inc.
- Date received
- 2025-06-13
- Decision date
- 2025-09-11
- Days to decision
- 90 days
- Decision code
- APPR
- Advisory committee
- Ophthalmic
- Expedited review
- No
- Location
- Irvine, CA
- Statement
- approval for a change in the endotoxin specification limit for preloaded intraocular lens storage/delivery systems to harmonize this limit with the relevant intraocular device ISO Standard and FDA Guidance.
Class III intelligence, same tools
PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.
510(k) clearances by this applicant
| Record | Firm | Date |
|---|---|---|
| ELITA™ Femtosecond Laser System, ELITA™ Patient Interface (K223566) | Johnson & Johnson Surgical Vision, Inc. | 2023-04-14 |
| VERITAS Phacoemulsification Console, VERITAS Advanced Fluidics Pack and Advanced Infusion Pack, VERITAS Swivel Handpiece, VERITAS Advanced Foot Pedal, VERITAS Remote Control (K203060) | Johnson & Johnson Surgical Vision, Inc. | 2021-03-12 |
| UNFOLDER Vitan Inserter (K191949) | Johnson & Johnson Surgical Vision, Inc. | 2019-09-13 |
| COMPACT INTUITIV System, COMPACT INTUITIV Wireless Remote Control, COMPACT INTUITIV Four-Button Foot Pedal, COMPACT INTUITIV Closed-Toe Foot Pedal (K191933) | Johnson & Johnson Surgical Vision, Inc. | 2019-08-16 |