Atlas FDA

SENSOLOG/DIALOG/REGENCY FAMILY OF PACEMAKERS

FDA premarket approval P880006/S055 · decided 2007-12-12

Approval details

PMA number
P880006
Supplement
S055
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Trade name
SENSOLOG/DIALOG/REGENCY FAMILY OF PACEMAKERS
Generic name
Pulse generator, permanent, implantable
Applicant
St. Jude Medical, Inc.
Date received
2007-10-24
Decision date
2007-12-12
Days to decision
49 days
Decision code
APPR
Product code
NVZ
Advisory committee
Cardiovascular
Expedited review
No
Location
Sylmar, CA
Statement
APPROVAL FOR TWO CORRECTIVE FIXES TO ADDRESS A HARDWARE ISSUED THAT CAN LEAVE THE CRT-D AND ICD SYSTEMS LISTED ABOVE IN A POST-SENSE REFRACTORY STATE. FOR DEVICES THAT ARE ALREADY IMPLANTED AND/OR ALREADY MANUFACTURED, CHANGES WILL BE MADE TO THE PROGRAMMER SOFTWARE. DURING A PATIENT FOLLOW-UP VISIT, THE PROGRAMMER WILL INTERROGATE THE DEVICE AND DOWNLOAD FIRMWARE RAM CODE TO CORRECT THE ANOMALOUS CONDITION IF IT OCCURS. FOR NEW DEVICES, A MODIFICATION HAS BEEN MADE TO THE SIO2 CHIP HARDWARE TO

Class III intelligence, same tools

PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.

510(k) clearances by this applicant

RecordFirmDate
Advisor HD Grid Mapping Catheter, Sensor Enabled (K172393)St. Jude Medical, Inc.2018-04-23
Confirm Rx Insertable Cardiac Monitor (ICM) System (K163407)St. Jude Medical, Inc.2017-09-29
PressureWire X Guidewire (K161171)St. Jude Medical, Inc.2016-08-15
Nanostim Introducer Kit (K161102)St. Jude Medical, Inc.2016-05-19
The Nanostim Introducer Kit (K160716)St. Jude Medical, Inc.2016-04-14
NANOSTIM INTRODUCER KIT (K140793)St. Jude Medical, Inc.2014-12-17
SJM CONFIRM IMPLANTABLE CARDIAC MONITOR (ICM) (K133481)St. Jude Medical, Inc.2014-05-28
ENLIGHTN RENAL GUIDE CATHETER (K131592)St. Jude Medical, Inc.2014-01-31