RESTORE ULTRA NEUROSTIMULATOR, RESTORE SENSOR NEUROSTIMULATOR, ITREL 4 NEUROSTIMULATOR AND ENTERNAL NEUROSTIMULATOR
FDA premarket approval P840001/S234 · decided 2013-06-19
Approval details
- PMA number
- P840001
- Supplement
- S234
- Supplement type
- Real-Time Process
- Supplement reason
- Change Design/Components/Specifications/Material
- Trade name
- RESTORE ULTRA NEUROSTIMULATOR, RESTORE SENSOR NEUROSTIMULATOR, ITREL 4 NEUROSTIMULATOR AND ENTERNAL NEUROSTIMULATOR
- Generic name
- Stimulator, spinal-cord, totally implanted for pain relief
- Applicant
- Medtronic Neuromodulation
- Date received
- 2013-02-07
- Decision date
- 2013-06-19
- Days to decision
- 132 days
- Decision code
- APPR
- Product code
- LGW
- Advisory committee
- Neurology
- Expedited review
- No
- Location
- Minneapolis, MN
- Statement
- APPROVAL FOR FIRMWARE CHANGES INTENDED TO CORRECT DEVICE BEHAVIORS THAT RESULT IN TEMPORARY LOSS OF STIMULATION (EXCEPTION BIT) AND CAN AFFECT PROGRAMMED STIMULATION PARAMETERS (OVER-DISCHARGE).