Atlas FDA

Restore, Itrel, Synergy and Intellis Spinal Cord Stimulation Systems and Pisces, Specify, and Vectris Spinal Cord Stimul

FDA premarket approval P840001/S468 · decided 2020-11-03

Approval details

PMA number
P840001
Supplement
S468
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Trade name
Restore, Itrel, Synergy and Intellis Spinal Cord Stimulation Systems and Pisces, Specify, and Vectris Spinal Cord Stimul
Generic name
Stimulator, spinal-cord, totally implanted for pain relief
Applicant
Medtronic Neuromodulation
Date received
2020-08-21
Decision date
2020-11-03
Days to decision
74 days
Decision code
APPR
Product code
LGW
Advisory committee
Neurology
Expedited review
No
Location
Minneapolis, MN
Statement
Approval to add the use of tin-lead plating for the terminations of both ends of the T8 tantalum capacitors and for the Model 97810 InterStim Micro INS only, change the solder paste stencil tooling for the 1411 capacitor size to increase the amount of solder paste to match the mount used on other capacitor case sizes.

Class III intelligence, same tools

PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.

510(k) clearances by this applicant

RecordFirmDate
SureTune4 Software (DEN210003)Medtronic Neuromodulation2021-08-23