Atlas FDA

MiniMed 670G System

FDA premarket approval P160017/S108 · decided 2023-04-18

Approval details

PMA number
P160017
Supplement
S108
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Trade name
MiniMed 670G System
Generic name
Automated insulin dosing device system, single hormonal control
Applicant
Medtronic Minimed, Inc.
Date received
2023-01-26
Decision date
2023-04-18
Days to decision
82 days
Decision code
APPR
Product code
OZP
Advisory committee
Clinical Chemistry
Expedited review
No
Location
Northridge, CA
Statement
Approval for a minor design change to the Next Generation Pumps (NGP) to remove remote bolus functionality, remove remote settings modification, and disable over the air programming. The proposed changes address vulnerabilities in connection with Field Action (FA1272) and will be implemented in the MiniMed 670G and 630G Insulin Pumps.

Class III intelligence, same tools

PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.

510(k) clearances by this applicant

RecordFirmDate
MiniMed Flex pump (K253743)Medtronic Minimed, Inc.2026-03-12
MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit (MMT-1894) (K253470)Medtronic Minimed, Inc.2026-01-15
SmartGuard technology; Predictive Low Glucose technology (K253585)Medtronic Minimed, Inc.2026-01-14
SmartGuard technology; Predictive Low Glucose technology (K251217)Medtronic Minimed, Inc.2025-08-29
MiniMed 780G insulin pump (K251032)Medtronic Minimed, Inc.2025-07-01
Extended Reservoir (K210714)Medtronic Minimed, Inc.2022-04-06
MiniMed Connect kit, MiniMed Connect uploader, MiniMed Connect app, CareLink Connect (CareLink Personal) (K151236)Medtronic Minimed, Inc.2015-05-19
MEDTRONIC MINIMED CARELINK USB CONNECTOR, MODEL NMT 7305 (K070438)Medtronic Minimed, Inc.2007-10-17