Atlas FDA

MEDTRONIC VIRTUOSO ICD (MODELS D154AWG AND D154VWC)

FDA premarket approval P980016/S102 · decided 2007-06-06

Approval details

PMA number
P980016
Supplement
S102
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Trade name
MEDTRONIC VIRTUOSO ICD (MODELS D154AWG AND D154VWC)
Generic name
Implantable cardioverter defibrillator (non-CRT)
Applicant
Medtronic Cardiac Rhythm Disease Management
Date received
2007-05-16
Decision date
2007-06-06
Days to decision
21 days
Decision code
APPR
Product code
LWS
Advisory committee
Cardiovascular
Expedited review
No
Location
Mounds View, MN
Statement
APPROVAL FOR A CHANGE IN THE BAROMETRIC PRESSURE REQUIREMENT FROM 6 ATMOSPHERES ABSOLUTE (ATA) TO 4 ATA AND A CHANGE IN BAROMETRIC PRESSURE TESTING METHODOLOGY USED DURING DESIGN VERIFICATION.