Atlas FDA

MEDTRONIC ICD'S

FDA premarket approval P980016/S115 · decided 2007-12-04

Approval details

PMA number
P980016
Supplement
S115
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Trade name
MEDTRONIC ICD'S
Generic name
Implantable cardioverter defibrillator (non-CRT)
Applicant
Medtronic Cardiac Rhythm Disease Management
Date received
2007-11-16
Decision date
2007-12-04
Days to decision
18 days
Decision code
APPR
Product code
LWS
Advisory committee
Cardiovascular
Expedited review
No
Location
Mounds View, MN
Statement
APPROVAL FOR MINOR DESIGN CHANGES THAT ADD TWO ALTERNATE TANTALUM CAPACITORS (XTC017, XTC018) WITH INCREASED VOLTAGE RATINGS TO APPROVED DEVICE COMPONENTS. THE DESIGN MODIFICATIONS ARE INTENDED TO REDUCE MANUFACTURING YIELD LOSS AND ASSOCIATED REWORK.

Class III intelligence, same tools