Atlas FDA

Master Restore, Itrel, Synergy, Intellis and Vanta Spinal Cord Stimulation Systems and Pisces, Specify, Vectris Spinal C

FDA premarket approval P840001/S498 · decided 2021-09-01

Approval details

PMA number
P840001
Supplement
S498
Supplement type
30-Day Notice
Supplement reason
Process Change - Manufacturer/Sterilizer/Packager/Supplier
Trade name
Master Restore, Itrel, Synergy, Intellis and Vanta Spinal Cord Stimulation Systems and Pisces, Specify, Vectris Spinal C
Generic name
Stimulator, spinal-cord, totally implanted for pain relief
Applicant
Medtronic Neuromodulation
Date received
2021-08-03
Decision date
2021-09-01
Days to decision
29 days
Decision code
OK30
Product code
LGW
Advisory committee
Neurology
Expedited review
No
Location
Minneapolis, MN
Statement
Approval for 1) equipment updates in the CEA 2-3 Silicone Molding area (e.g., six additional ovens), updated option for use of temperature recorder (Vaisala Monitoring System Model VL-1700-54T), updated oven mode for the new ovens to use Manual Mode, removal of monitor lot for all ovens, removal of setup lot of new ovens, addition of a new and smaller basket for oven use, and addition of a new toolmaker scope (Nikon MM-40), 2) an update to the applicable work instructions (WI); and 3) the oven o

Class III intelligence, same tools

PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.

510(k) clearances by this applicant

RecordFirmDate
SureTune4 Software (DEN210003)Medtronic Neuromodulation2021-08-23