IVT CORONARY ATHERECTOMY SYSTEM (TEC)(R)
FDA premarket approval P910019/S003 · decided 1998-01-29
Approval details
- PMA number
- P910019
- Supplement
- S003
- Supplement type
- Special (Immediate Track)
- Supplement reason
- Process Change - Manufacturer/Sterilizer/Packager/Supplier
- Trade name
- IVT CORONARY ATHERECTOMY SYSTEM (TEC)(R)
- Generic name
- CATHETER, CORONARY, ATHERECTOMY
- Applicant
- Intensive Technology, Inc.
- Date received
- 1997-10-28
- Decision date
- 1998-01-29
- Days to decision
- 93 days
- Decision code
- APPR
- Product code
- MCX
- Advisory committee
- Cardiovascular
- Expedited review
- No
- Location
- San Diego, CA
- Statement
- Approval for a change to the 5.5F TEC(R) cutter catheter, involving an increase in the skirt length of the cutter component from 0.079 inches to 0.104 inches.
Class III intelligence, same tools
PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.
510(k) clearances by this applicant
| Record | Firm | Date |
|---|---|---|
| INTRA-OPERATIVE TEC ATHERECTOMY SYSTEM (K922970) | Intensive Technology, Inc. | 1993-01-13 |
| TEC ROTATING DUAL HEMOVALVE (K912664) | Intensive Technology, Inc. | 1992-08-24 |
| IVT GUIDING CATHETER (K911869) | Intensive Technology, Inc. | 1991-10-16 |
| IVT INTRODUCER SHEATH (K903463) | Intensive Technology, Inc. | 1991-01-03 |
| HEMOCHRON(R) HEPARIN RESPONSE TEST (K904003) | Intensive Technology, Inc. | 1990-11-29 |
| MODIFIED IVT PERIPHERAL ATHERECTOMY (K901767) | Intensive Technology, Inc. | 1990-06-28 |
| MODIFIED IVT PERIPHERAL ATHERECTOMY SYSTEM (K896403) | Intensive Technology, Inc. | 1990-03-06 |
| IVT 0.014 GUIDE WIRE (K895903) | Intensive Technology, Inc. | 1990-01-04 |