HYDRO THERMABLATOR (HTA) ENDOMETRIAL ABLATION SYSTEM
FDA premarket approval P000040/S008 · decided 2006-06-23
Approval details
- PMA number
- P000040
- Supplement
- S008
- Supplement type
- Real-Time Process
- Supplement reason
- Change Design/Components/Specifications/Material
- Trade name
- HYDRO THERMABLATOR (HTA) ENDOMETRIAL ABLATION SYSTEM
- Generic name
- Device, thermal ablation, endometrial
- Applicant
- Axora Medical, Inc.
- Date received
- 2006-05-15
- Decision date
- 2006-06-23
- Days to decision
- 39 days
- Decision code
- APPR
- Product code
- MNB
- Advisory committee
- Obstetrics/Gynecology
- Expedited review
- No
- Location
- Jacksonville, FL
- Statement
- APPROVAL FOR 1) ADDITION OF THE LINEAR FLASH CARD MEMORY TO THE CENTRAL PROCESSING BOARD; 2) TEMPERATURE CONTROLLER PID SETTINGS/INCREASE IN CYCLE TIME PROGRAMMED INTO HEATER LOGIC AND CONTROLLER BOARD; AND 3) ADDITION OF COMPONENTS TO THE LEVEL SENSOR BOARD.
Class III intelligence, same tools
PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.