HIRESOLUTION BIONIC EAR SYSTEM
FDA premarket approval P960058/S037 · decided 2004-07-23
Approval details
- PMA number
- P960058
- Supplement
- S037
- Supplement type
- Special (Immediate Track)
- Trade name
- HIRESOLUTION BIONIC EAR SYSTEM
- Generic name
- Implant, cochlear
- Applicant
- Advanced Bionics
- Date received
- 2004-05-10
- Decision date
- 2004-07-23
- Days to decision
- 74 days
- Decision code
- APPR
- Product code
- MCM
- Advisory committee
- Ear, Nose, Throat
- Expedited review
- No
- Location
- Valencia, CA
- Statement
- APPROVAL FOR EXPANSION OF THE IN-PROCESS INSPECTION PROCEDURE FOR THE HIRES 90K IMPLANT. NAMELY, THE PMA SUPPLEMENT PROPOSED TO ADD AN ADDITIONAL VISUAL INSPECTION STEP TO ENSURE WELD INTEGRITY BY USING A MINIMUM OF 80X MAGNIFICATION TO VERIFY THE FOLLOWING: 1) COMPLETE CLOSURE OF THE WELD HOLD, 2) ACCEPTABLE WELD QUALITY, AND 3) ABSENCE OF FOREIGN MATERIAL AT THE WELD SITE.
Class III intelligence, same tools
PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.