Atlas FDA

HIRESOLUTION BIONIC EAR SYSTEM

FDA premarket approval P960058/S037 · decided 2004-07-23

Approval details

PMA number
P960058
Supplement
S037
Supplement type
Special (Immediate Track)
Trade name
HIRESOLUTION BIONIC EAR SYSTEM
Generic name
Implant, cochlear
Applicant
Advanced Bionics
Date received
2004-05-10
Decision date
2004-07-23
Days to decision
74 days
Decision code
APPR
Product code
MCM
Advisory committee
Ear, Nose, Throat
Expedited review
No
Location
Valencia, CA
Statement
APPROVAL FOR EXPANSION OF THE IN-PROCESS INSPECTION PROCEDURE FOR THE HIRES 90K IMPLANT. NAMELY, THE PMA SUPPLEMENT PROPOSED TO ADD AN ADDITIONAL VISUAL INSPECTION STEP TO ENSURE WELD INTEGRITY BY USING A MINIMUM OF 80X MAGNIFICATION TO VERIFY THE FOLLOWING: 1) COMPLETE CLOSURE OF THE WELD HOLD, 2) ACCEPTABLE WELD QUALITY, AND 3) ABSENCE OF FOREIGN MATERIAL AT THE WELD SITE.

Class III intelligence, same tools

PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.