HELISTAT/HELITENE ABSORBABLE COLLAGEN HEMOSTATIC AGENTS
FDA premarket approval P850010/S053 · decided 2014-10-17
Approval details
- PMA number
- P850010
- Supplement
- S053
- Supplement type
- 135 Review Track For 30-Day Notice
- Supplement reason
- Process Change - Manufacturer/Sterilizer/Packager/Supplier
- Trade name
- HELISTAT/HELITENE ABSORBABLE COLLAGEN HEMOSTATIC AGENTS
- Generic name
- Agent, absorbable hemostatic, collagen based
- Applicant
- Integra LifeSciences Corporation
- Date received
- 2014-04-11
- Decision date
- 2014-10-17
- Days to decision
- 189 days
- Decision code
- APPR
- Product code
- LMF
- Advisory committee
- General, Plastic Surgery
- Expedited review
- No
- Location
- Plainsboro, NJ
- Statement
- APPROVAL FOR THE IMPLEMENTATION OF MECHANICAL, PROGRAM, AND SYSTEM CONTROL CHANGES TO THE EXISTING CHILLED WATER SYSTEM THAT SERVICES COLLAGEN LYOPHILIZER MACHINES IN THE LYO MANUFACTURING AREA OF BUILDING 1 AT THE INTEGRA NEUROSCIENCES FACILITY IN AÑASCO, PUERTO RICO, FOR THE DEVICES.
Class III intelligence, same tools