Gore Excluder Conformable AAA Endoprosthesis (CEXC)

FDA premarket approval P200030 · decided 2020-12-22

Approval details

PMA number
P200030
Trade name
Gore Excluder Conformable AAA Endoprosthesis (CEXC)
Generic name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Applicant
W. L. Gore and Associates, Inc.
Date received
2020-06-29
Decision date
2020-12-22
Days to decision
176 days
Decision code
APPR
Product code
MIH
Advisory committee
Cardiovascular
Expedited review
No
Location
Phonix, AZ
Statement
Approval for Gore Excluder Conformable AAA Endoprosthesis (EXCC)This device is indicated to exclude the aneurysm from the blood circulation in patients diagnosed with infrarenal abdominal aortic aneurysm (AAA) disease and who have appropriate anatomy as described below:1) Adequate iliac / femoral access;2) Infrarenal aortic neck treatment diameter range of 16–32 mm;3) A minimum aortic neck length of 15 mm when proximal aortic neck angulation is =<60 degrees; and4) Iliac artery treatment diameter