Atlas FDA

ELA RATE-RESPONSIVE PACEMAKERS

FDA premarket approval P950029/S011 · decided 2000-10-11

Approval details

PMA number
P950029
Supplement
S011
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Trade name
ELA RATE-RESPONSIVE PACEMAKERS
Generic name
implantable pacemaker Pulse-generator
Applicant
MicroPort CRM USA, Inc.
Date received
2000-08-18
Decision date
2000-10-11
Days to decision
54 days
Decision code
APPR
Product code
DXY
Advisory committee
Cardiovascular
Expedited review
No
Location
Arlington, TN
Statement
APPROVAL IS FOR THE FOLLOWING MODIFICATIONS TO YOUR BRIO FAMILY OF PACEMAKERS: 1) MODIFY THE VALUE FOR THE NUMBER OF CYCLES THE DEVICE WILL ALLOW BEFORE A SELF TEST IS INITIATED WHEN THE SENSOR "CALLS FOR AN ELEVATED PACING RATE FOR A DURATION OR FREQUENCY THAT EXCEES REASONABLE PHYSIOLOGIC LIMITS" FROM 30,000 TO 100,000 CYCLES; AND2) CHANGING THE MINIMUM SENSING THRESHOLD DURING T-WAVE PROTECTION FROM 2.2 TO 2.5MV.