DUETT PRO SEALING DEVICE AND THE DIAGNOSTIC DUETT PRO SEALING DEVICE SEALING DEVICE

FDA premarket approval P990037/S012 · decided 2003-01-30

Approval details

PMA number
P990037
Supplement
S012
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Trade name
DUETT PRO SEALING DEVICE AND THE DIAGNOSTIC DUETT PRO SEALING DEVICE SEALING DEVICE
Generic name
Device, hemostasis, vascular
Applicant
Vascular Solutions, Inc.
Date received
2003-01-17
Decision date
2003-01-30
Days to decision
13 days
Decision code
APPR
Product code
MGB
Advisory committee
Cardiovascular
Expedited review
No
Location
Minneapolis, MN
Statement
APPROVAL FOR A DESIGN CHANGE TO THE DUETT PRO SEALING DEVICE AND THE DIAGNOSTIC DUETT PRO SEALING DEVICE. SPECIFICALLY, 1) REMOVAL OF THE SILICONE GASKET FROM THE PRESSURE RELIEF VALVE DESIGN; 2) REMOVAL OF A LUER LOCK THREAD/ATTACHMENT WHICH WAS USED TO CONNECT THE CATHETER TO THE PRESSURE RELIEF VALVE AND REPLACE WITH AN ADHESIVE BOND; 3) REMOVAL OF THE CAP COVER AND THREADS FROM THE RELIEF VALVE, AND 4) SHORTENING OF THE LENGTH AND HEIGHT OF THE RELIEF VALVE.

Class III intelligence, same tools

PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.

510(k) clearances by this applicant

RecordFirmDate
Octane Mechanical Thrombectomy System (K182232)Vascular Solutions, Inc.2018-09-11
Bandit guidewire (K181647)Vascular Solutions, Inc.2018-07-20
VSI Micro-Introducer Kit (K180913)Vascular Solutions, Inc.2018-05-09
TrapLiner catheter (K180088)Vascular Solutions, Inc.2018-04-04
Warrior 14 guidewire (K180128)Vascular Solutions, Inc.2018-02-16
Gel-Bead (600-800 micron) - 1 ml sphere volume, Gel-Bead (600-800 micron) - 2 ml sphere volume (K173891)Vascular Solutions, Inc.2018-01-19
Raider Guidewire (K173532)Vascular Solutions, Inc.2017-12-15
Octane aspiration system (K173266)Vascular Solutions, Inc.2017-12-01