Atlas FDA

DROMOS DR/SR RATE ADAPTIVE PACING SYSTEM

FDA premarket approval P950037/S108 · decided 2012-05-17

Approval details

PMA number
P950037
Supplement
S108
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Trade name
DROMOS DR/SR RATE ADAPTIVE PACING SYSTEM
Generic name
Pulse generator, permanent, implantable
Applicant
Biotronik, Inc.
Date received
2012-04-05
Decision date
2012-05-17
Days to decision
42 days
Decision code
APPR
Product code
NVZ
Advisory committee
Cardiovascular
Expedited review
No
Location
Lake Oswego, OR
Statement
APPROVAL TO ADD A COPPER ALLOY SHIELDING PLATE TO BE APPLIED OVER THE HOUSING OF A QUARTZ CRYSTAL ON THE PCB AND UPDATE THE DEVICE LABELING. THE PURPOSE FOR THESE MODIFICATIONS IS TO IMPROVE PERFORMANCE DURING RF CAUTERY.

Class III intelligence, same tools

PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.

510(k) clearances by this applicant

RecordFirmDate
BIOMONITOR IV (471155) (K261074)Biotronik, Inc.2026-05-01
Passeo-35 Xeo Peripheral Dilatation Catheter; Passeo-18 Peripheral Dilatation Catheters; Passeo-14 Peripheral Dilatation Catheter; Oscar Peripheral Multifunctional Catheter System; Pantera Pro Percutaneous Transluminal Coronary Angioplasty Catheter; Pantera LEO Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter (K250706)Biotronik, Inc.2025-04-04
Pantera Pro; Pantera LEO (K242969)Biotronik, Inc.2024-10-25
Selectra 3D Lead Delivery System (443624-443629, 451789-451791); Selectra Slitter Tool (383119); Selectra Accessory Kit (375518) (K240787)Biotronik, Inc.2024-09-12
Oscar Peripheral Multifunctional Catheter system (K241711)Biotronik, Inc.2024-07-11
BIOMONITOR IV (K230375)Biotronik, Inc.2023-05-19
Passeo-35 Xeo Peripheral Dilatation Catheter (K222065)Biotronik, Inc.2023-02-16
BIOMONITOR IIIm, BIOMONITOR III (K221856)Biotronik, Inc.2022-07-27