CryoConsole
FDA premarket approval P020045/S104 · decided 2023-05-31
Approval details
- PMA number
- P020045
- Supplement
- S104
- Supplement type
- Real-Time Process
- Supplement reason
- Change Design/Components/Specifications/Material
- Trade name
- CryoConsole
- Generic name
- Cardiac ablation percutaneous catheter
- Applicant
- Medtronic Cryocath, LP
- Date received
- 2023-03-29
- Decision date
- 2023-05-31
- Days to decision
- 63 days
- Decision code
- APPR
- Product code
- LPB
- Advisory committee
- Cardiovascular
- Expedited review
- No
- Location
- Mounds View, MN
- Statement
- Approval for an additional CryoConsole configuration (CFN 106E2) that accepts 220V input power and includes additional components to limit peak current during power-up.
Class III intelligence, same tools
PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.
510(k) clearances by this applicant
| Record | Firm | Date |
|---|---|---|
| FlexCath Advance Steerable Sheath and Dilator (K202620) | Medtronic Cryocath, LP | 2020-11-13 |
| FlexCath Advance Steerable Sheath and Dilator (K183174) | Medtronic Cryocath, LP | 2018-11-26 |
| FlexCath Select Steerable Shealth and Dilator (K163268) | Medtronic Cryocath, LP | 2017-02-02 |
| FlexCath Select Steerable Sheath (10 French) (K142684) | Medtronic Cryocath, LP | 2014-11-06 |
| FLEXCATH ADVANCE STEERABLE SHEATH (12 FRENCH) (K123591) | Medtronic Cryocath, LP | 2012-12-27 |
| FLEXCATH STEERABLE SHEATH (12 FRENCH), FLEXCATH STEERABLE SHEATH (10 FRENCH) MODEL 3FC12, 3FC10 (K102176) | Medtronic Cryocath, LP | 2010-09-01 |