Atlas FDA

ANIMAS VIBE SYSTEM

FDA premarket approval P130007/S019 · decided 2016-11-05

Approval details

PMA number
P130007
Supplement
S019
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Trade name
ANIMAS VIBE SYSTEM
Generic name
Pump, infusion, insulin, to be used with invasive glucose sensor
Applicant
Animas Corp.
Date received
2016-07-29
Decision date
2016-11-05
Days to decision
99 days
Decision code
APPR
Product code
OYC
Advisory committee
Clinical Chemistry
Expedited review
No
Location
Wayne, PA
Statement
Approval for modifications to the attachment method that secures the disposable housing to the adhesive patch of the sensor pod of Dexcom’s G4 PLATINUM/G5 Mobile sensor component, which required modifications to the disposable housing, implementation of a semi-automated heat staking machine and a change to the inspection method. The sensor is a component of the Animas Vibe System.

Class III intelligence, same tools

PMAs share the API, exports and alerts with 510(k) data - watch an applicant and hear about every approval.

510(k) clearances by this applicant

RecordFirmDate
SYMPHONY GLUCOSE MANAGEMENT SYSTEM (K080639)Animas Corp.2008-06-24
EZ MANAGER MAX DIABETES MANAGEMENT SOFTWARE (K080587)Animas Corp.2008-05-15
EZMANAGER MAX DIABETES MANAGEMENT SOFTWARE (K063674)Animas Corp.2007-03-30
ANIMAS MODEL IR1250 INSULIN INFUSION PUMP (K042873)Animas Corp.2004-12-10
MODEL IR 1200 INSULIN PUMP (K032257)Animas Corp.2003-10-16
EZMANAGER PLUS (K022394)Animas Corp.2003-06-05
ANIMUS EZSERTER INFUSION SET INSERTER (K021700)Animas Corp.2002-06-04
MODEL R1000 IR INSULIN PUMP, EZLINK SOFTWARE, SERMON SOFTWARE, R1000 IR DOWNLOAD CRADLE, AND SERIAL INTERFACE ADAPTOR DO (K021439)Animas Corp.2002-05-29